Draegar Medical Systems, Inc. Infinity M300 telemetry monitoring device Software versions VG2.3.1 and lower, Model Numbers MS25755, MS26076, MS26031, MS18623, MS22862, MS25301 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Infinity M300 telemetry monitoring device Software versions VG2.3.1 and lower, Model Numbers MS25755, MS26076, MS26031, MS18623, MS22862, MS25301
Brand
Draegar Medical Systems, Inc.
Lot Codes / Batch Numbers
All systems using Software version VG2.3.1 and lower
Products Sold
All systems using Software version VG2.3.1 and lower
Draegar Medical Systems, Inc. is recalling Infinity M300 telemetry monitoring device Software versions VG2.3.1 and lower, Model Numbers MS25755 due to The devices have potential cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDoS), Spoofing, and Tampering.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The devices have potential cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDoS), Spoofing, and Tampering.
Recommended Action
Per FDA guidance
An Urgent Medical Device Recall notification letter dated 10/9/19 was sent to customers. Recommended cybersecurity precautions: While we are in the process of updating the software, we recommend that you follow best recommended practices to limit access to the devices and wireless network: " Physical security of the patient monitors is recommended and is the responsibility of the operating organization. " Physical security boundaries to prevent access to your facility's network infrastructure is recommended and is the responsibility of the operating organization. " Drager relies on the medical device isolation mechanism of the VLANs and the proper configuration, implementation, and use of the operating organization's security measures to prevent the introduction of malware onto the Infinity Network. Action required: Please complete the attached Acknowledgement and Response Form and return it to Drager per instructions included on the form. The software upgrade will be provided free of charge. The upgrade path for your M300 device is dependent on the part number of your M300s currently in use. Your local Drager representative will contact you prior to the availability of software version VG2.3.2 to discuss the upgrade path. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM - 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.