Draeger Medical, Inc. Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use.
Brand
Draeger Medical, Inc.
Lot Codes / Batch Numbers
Catalog Numbers 5704757 and 5704758.
Products Sold
Catalog Numbers 5704757 and 5704758.
Draeger Medical, Inc. is recalling Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use. due to The device may not generate a pilot line disconnect alarm if the pilot line becomes disconnected, which may potentially result in a substantial reduct. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The device may not generate a pilot line disconnect alarm if the pilot line becomes disconnected, which may potentially result in a substantial reduction in tidal volume delivery.
Recommended Action
Per FDA guidance
An "Urgent - Medical Device Recall" Letter dated January 2009 was issued to consignees via certified mail. The letter described the issue, specific conditions, background information, and precautions for customers. An additional letter dated January 22, 2009 was included with the "Urgent - Medical Device Recall" letter which instructed consignees to inform an enduser of the contents of the notification letter because some of the consignees are home care device/service providers. When the new hardware/software solution is available, affected devices will be updated free of charge. Direct questions about the notification letter to Drager Medical, Inc. by calling 1-800-543-5047 (press "1" at the prompt and then press "2349#"). Direct questions regarding updating your device when the new software is available, contact Drager Service Technical Support at 1-800-543-5047 (press "4" at the prompt).
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026