Drew Scientific Group PLC Bayer Advia 70 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Bayer Advia 70 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX.
Brand
Drew Scientific Group PLC
Lot Codes / Batch Numbers
All lots.
Products Sold
All lots.
Drew Scientific Group PLC is recalling Bayer Advia 70 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX. due to System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts.
Recommended Action
Per FDA guidance
Firm sent 'Technical Bulletin #39' to all customers on 09/20/06. Users advised of potential for inaccurate readings and provided with instructions on how to eliminate problem.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
FL, MA, MN, NM, NY, ND, TX, PR