DWL Systems, Inc. EZ-DOP, diagnostic ultrasound imaging system Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
EZ-DOP, diagnostic ultrasound imaging system
Brand
DWL Systems, Inc.
Lot Codes / Batch Numbers
Serial numbers LDP4 0024 through LDP4 390.
Products Sold
Serial numbers LDP4 0024 through LDP4 390.
DWL Systems, Inc. is recalling EZ-DOP, diagnostic ultrasound imaging system due to A certain sequence of data entries on the keypad could cause the ultrasound medical device to display incorrect transmission power.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A certain sequence of data entries on the keypad could cause the ultrasound medical device to display incorrect transmission power.
Recommended Action
Per FDA guidance
The firm notified consignees by mail on 1/5/2004 and advised of potential for inaccurate power display when the device is actually transmitting. Consignees were requested to return the device for a software upgrade.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026