Dynex Technologies, Inc. Pipette Sample Tip, Product part #65910, packaged in cardboard cartons containing 4 plastic racks of 108 pipette tips per rack. Pipette tips are accessory component for Dynex Technologies proprietary automated DSX system. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Pipette Sample Tip, Product part #65910, packaged in cardboard cartons containing 4 plastic racks of 108 pipette tips per rack. Pipette tips are accessory component for Dynex Technologies proprietary automated DSX system.
Brand
Dynex Technologies, Inc.
Lot Codes / Batch Numbers
Lot 412410 printed on carton
Products Sold
Lot 412410 printed on carton
Dynex Technologies, Inc. is recalling Pipette Sample Tip, Product part #65910, packaged in cardboard cartons containing 4 plastic racks of due to An accessory sampling tip for an in vitro diagnostic test instrument may be inadequately sealed and may cause false assay patient results.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
An accessory sampling tip for an in vitro diagnostic test instrument may be inadequately sealed and may cause false assay patient results.
Recommended Action
Per FDA guidance
The recalling firm notified consignees by faxed letter on 6/11/04 and follow-up letter by mail 6/24/04. The notice advised of vacuum failure associated with lot 412410 of sample tips and recommended quarantine of the defective tips. Consignees were requested to return response form. The 6/24/04 notification recommends that consignees discard/destroy lot 412410 and asks distributors to notify their customers. The firm is working with supplier of the pipette tips to replace recalled product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution