E-Z-EM, Inc Brand Name: EmpowerCT/EmpowerCTA Injector System Ceiling Column Accessory; Common/Usual Name: CT Injector System Ceiling Column Accessory, Catalog Number 9886S. Ceiling Column Accessory: The CT and CTA Injector Systems employs an Articulating Arm that mounts to the Ceiling Column Accessory, which are designated as 9886S (11.25''), M (21.50''), and L (34.50''), i.e. 9886L. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Brand Name: EmpowerCT/EmpowerCTA Injector System Ceiling Column Accessory; Common/Usual Name: CT Injector System Ceiling Column Accessory, Catalog Number 9886S. Ceiling Column Accessory: The CT and CTA Injector Systems employs an Articulating Arm that mounts to the Ceiling Column Accessory, which are designated as 9886S (11.25''), M (21.50''), and L (34.50''), i.e. 9886L.
Brand
E-Z-EM, Inc
Lot Codes / Batch Numbers
No lot #s or serial #s assigned to this accessory
Products Sold
No lot #s or serial #s assigned to this accessory
E-Z-EM, Inc is recalling Brand Name: EmpowerCT/EmpowerCTA Injector System Ceiling Column Accessory; Common/Usual Name: CT Inj due to Out of specification welding could cause the ceiling mount accessory to detach from its mounting. The firm received 1 complaint from a hospital that t. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Out of specification welding could cause the ceiling mount accessory to detach from its mounting. The firm received 1 complaint from a hospital that the injector fell off the ceiling mounting bracket and landed on the floor.
Recommended Action
Per FDA guidance
On 12/3/04, "Urgent: Medical Device Recall" Letters (dated 12/3/04), were mailed via return receipt.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026