ela Medical Llc ELA Rhapsody SR model 2210. A single chamber rate-responsive pacemaker that uses both minute ventilation and accelerometer sensors. Sterilization with Ethylene Oxide Gas. Sorin Group, ELA medical, 98, rue Maurice Amoux, 92120 Montrouge - France. Made in France. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ELA Rhapsody SR model 2210. A single chamber rate-responsive pacemaker that uses both minute ventilation and accelerometer sensors. Sterilization with Ethylene Oxide Gas. Sorin Group, ELA medical, 98, rue Maurice Amoux, 92120 Montrouge - France. Made in France.
Brand
ela Medical Llc
Lot Codes / Batch Numbers
Rhapsody SR 2210 Serial numbers: 244WE078, 249WE059, 309WE066, 309WE145, 316WE215, 316WE233 AND 321WE080.
Products Sold
Rhapsody SR 2210 Serial numbers: 244WE078, 249WE059, 309WE066, 309WE145, 316WE215, 316WE233 AND 321WE080.
ela Medical Llc is recalling ELA Rhapsody SR model 2210. A single chamber rate-responsive pacemaker that uses both minute ventil due to Through post-market surveillance, ELA Medical has observed that a no-output condition could occur in a limited number of Symphony or rhapsody pacemake. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Through post-market surveillance, ELA Medical has observed that a no-output condition could occur in a limited number of Symphony or rhapsody pacemakers. The no-output condition could occur due to metal migration caused by a specific manufacturing process.
Recommended Action
Per FDA guidance
A Dear Doctor letter, dated 10/25/05, was sent to all affected doctors. The letter describes the failures and mentions that pacemaker depend patients implanted with devices manufactured in the first group could require prophylactic replacement.