Enpath Medical, Inc Angel Medical Systems Lead Adaptor system adapts a St. Jude Model 1488T Bipolar Endocardial Lead to an AngelMed Guardian IMD. Manufactured for Angel Medical Systems, Inc. 1 Shelia Drive, Tinton Falls, NJ; Part No. 0208-2100-001, provided to Angel Medical Systems for International Clinical Studies; Sterile/ EO, Single Use. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Angel Medical Systems Lead Adaptor system adapts a St. Jude Model 1488T Bipolar Endocardial Lead to an AngelMed Guardian IMD. Manufactured for Angel Medical Systems, Inc. 1 Shelia Drive, Tinton Falls, NJ; Part No. 0208-2100-001, provided to Angel Medical Systems for International Clinical Studies; Sterile/ EO, Single Use.
Brand
Enpath Medical, Inc
Lot Codes / Batch Numbers
Model LA-001, Lot # W38932
Products Sold
Model LA-001, Lot # W38932
Enpath Medical, Inc is recalling Angel Medical Systems Lead Adaptor system adapts a St. Jude Model 1488T Bipolar Endocardial Lead to due to The internal component (Connector Block) of the Lead Adapter used to secure a lead to the adapter has exhibited varying degress of radial cracks that . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The internal component (Connector Block) of the Lead Adapter used to secure a lead to the adapter has exhibited varying degress of radial cracks that could cause a temporary intermittent electrical signal disruption if lateral tension is applied to device.
Recommended Action
Per FDA guidance
On July 7th, Enpath determined that a recall of Model 501206 and 501214 Adapters should be initiated and contacted customer via telephone and formally notified via letter on July 10, 2006. On July 14th, Enpath determined that a recall of Model 501204, Model 501205 and Angel Medical Model LA-001 Adapters should be initiated and two more consignees were notified via telephone on July 17th and both formally notified via letter on July 18, 2006. Enpath requested that customers segregate and return any product from the lot numbers listed in the letter, cease distribution of these devices and notify customers.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, MN, NJ