Essential Medical Supply, Inc. ESSENTIAL MEDICAL SUPPLY, INC. ENDURANCE HD, BUILT TO LAST, HEAVY DUTY FOUR WHEEL WALKER for users up to 500 pounds. Model W1801B- Blue, UPC 7 54756 91802 6 all serial numbers. W1801R-Red, UPC 7 54756 91803 3 all serial numbers. The Endurance HD is a four wheeled steel frame walker with loop locking wheel locks, a padded seat and an under seat tote for carrying personal articles. Used to assist in ambulation and specifically for obese patients weighing up to 500 pounds. Warranty Card: ".. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ESSENTIAL MEDICAL SUPPLY, INC. ENDURANCE HD, BUILT TO LAST, HEAVY DUTY FOUR WHEEL WALKER for users up to 500 pounds. Model W1801B- Blue, UPC 7 54756 91802 6 all serial numbers. W1801R-Red, UPC 7 54756 91803 3 all serial numbers. The Endurance HD is a four wheeled steel frame walker with loop locking wheel locks, a padded seat and an under seat tote for carrying personal articles. Used to assist in ambulation and specifically for obese patients weighing up to 500 pounds. Warranty Card: "..
Brand
Essential Medical Supply, Inc.
Lot Codes / Batch Numbers
All SS Numbers 05F0002740 to RPN090702497.
Products Sold
All SS Numbers 05F0002740 to RPN090702497.
Essential Medical Supply, Inc. is recalling ESSENTIAL MEDICAL SUPPLY, INC. ENDURANCE HD, BUILT TO LAST, HEAVY DUTY FOUR WHEEL WALKER for users u due to Essential Medical Supply, Inc. is recalling W1801B and W1801R Endurance HD Walkers due to recent reports and tests that show the front caster may deta. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Essential Medical Supply, Inc. is recalling W1801B and W1801R Endurance HD Walkers due to recent reports and tests that show the front caster may detach from the frame. The potential failures include a collapse of the plastic fork or shearing of the stud attaching the caster to the frame.
Recommended Action
Per FDA guidance
All Consignees were mailed an Urgent: Product Recall W1801B and W1801R Endurance HD Walkers notification on December 30, 2009. All notifications were sent first class mail in a 9" X 6" envelope. Response cards were included. The letter identified the affected product and the reason for recall. Customers were asked to examine their inventory for affected product and to discontinue further use or distribution of the product. If the product was further distributed, customers should contact their customers, advise them of the recall and have them return the units to them. Customers should follow the instructions provided for return and replacement of units.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026