Ethicon Endo-Surgery FlexTrayTM Endopath Gen. Laparoscopy Tray, Kit Code TGL45. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
FlexTrayTM Endopath Gen. Laparoscopy Tray, Kit Code TGL45.
Brand
Ethicon Endo-Surgery
Lot Codes / Batch Numbers
V41Y2R, V42A66, V42E3C, V42P37, V42R52, V4328E.
Products Sold
V41Y2R, V42A66, V42E3C, V42P37, V42R52, V4328E.
Ethicon Endo-Surgery is recalling FlexTrayTM Endopath Gen. Laparoscopy Tray, Kit Code TGL45. due to Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.
Recommended Action
Per FDA guidance
The recalling firm sent letters to hospitals and distributors, dated 12/14/04.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026