Ethicon Endo-Surgery Inc ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45; ATB45; ATW45; ATS45NK; and NAW45. Intended for transection, resection, and/or creation of anastomoses. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45; ATB45; ATW45; ATS45NK; and NAW45. Intended for transection, resection, and/or creation of anastomoses.
Brand
Ethicon Endo-Surgery Inc
Lot Codes / Batch Numbers
The recalled Lot Numbers are: E4MJ47, E4MF6W, E4MK84, E4MM23, E4MM64, E4MP4A, E4MR15, E4MT07, E4MT4Y, F4MU2R, F4MY91, E4MG53, E4MG5K, E4MG5L, E4MN39, E4MG89, E4MG9M, E4MH2L, E4ML6J, E4MG77, E4MG8A, E4MH18, E4MJ3K, F4MX8C, and F4N799.
Products Sold
The recalled Lot Numbers are: E4MJ47, E4MF6W, E4MK84, E4MM23, E4MM64, E4MP4A, E4MR15, E4MT07, E4MT4Y, F4MU2R, F4MY91, E4MG53, E4MG5K, E4MG5L, E4MN39, E4MG89, E4MG9M, E4MH2L, E4ML6J, E4MG77, E4MG8A, E4MH18, E4MJ3K, F4MX8C, and F4N799.
Ethicon Endo-Surgery Inc is recalling ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; A due to The firm initiated this second subsequent recall because the affected devices can remain closed and clamped on tissue during surgical procedures due t. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The firm initiated this second subsequent recall because the affected devices can remain closed and clamped on tissue during surgical procedures due to the fact that the mechanism which connects the articulation joint to the device shaft is insufficient.
Recommended Action
Per FDA guidance
Ethicon Endo-Surgery issued "Urgent Expanded Device Recall Event 1981" notification letters dated April 7, 2009, to their customers. They provided a list of all Affected Product Codes of the ETS FLEX45 Articulating Endoscopic Linear Cutters (including the Product Code: ATS45 from the 1st recall) and Affected Flex Tray Codes subject to recall. The notification document reminds the Operating Room and Materials Management customers of the problem which necessitated the previous recall (Z-0361-2011). It goes on to instruct their customers to immediately discontinue use of the affected products. They are to complete and fax back the Business Reply Form, Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle, Replacement product should be received within 30-60 days after return, Customers can contact Stericycle concerning this recall at 1-888-365-4680, reference Event 1981.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.