Exactech, Inc. ARTHROFOCUS Polyethylene Tibial Inserts (Product Line: 256-12-XX, 256-13-XX). NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ARTHROFOCUS Polyethylene Tibial Inserts (Product Line: 256-12-XX, 256-13-XX). NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.
Brand
Exactech, Inc.
Lot Codes / Batch Numbers
Item Number (Device Identifier, Serial number(s)): 256-12-02 (No Device Identifier, 1208624, 1208625, 1208626, 1208627), 256-12-03 (10885862066633, 1208774, 1446326), 256-13-02 (No Device Identifier, 1208710, 1208715), 256-13-03 (No Device Identifier, 1208859, 1208863), 256-13-04 (No Device Identifier, 1209010, 1209013)
Products Sold
Item Number (Device Identifier; Serial number(s)): 256-12-02 (No Device Identifier; 1208624, 1208625, 1208626, 1208627); 256-12-03 (10885862066633; 1208774, 1446326); 256-13-02 (No Device Identifier; 1208710, 1208715); 256-13-03 (No Device Identifier; 1208859, 1208863); 256-13-04 (No Device Identifier; 1209010, 1209013)
Exactech, Inc. is recalling ARTHROFOCUS Polyethylene Tibial Inserts (Product Line: 256-12-XX, 256-13-XX). NOTE: Information r due to Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Recommended Action
Per FDA guidance
Exactech notified distributors and sales representatives on about 08/30/2021 via letter titled "URGENT MEDICAL DEVICE RECALL." Actions included removing all Knee and Ankle UHMWPE products labeled with an 8-year shelf life and not packaged in EVOH/Nylon bags, in a phased approach over 12 months. Phase 1: immediately return all knee and ankle UHMWPE devices labeled with an 8-year shelf life that will be 5 years old or older by 08/31/2022 not packaged in EVOH/Nylon bags. Phase 2: between 05/31/2022 to 08/31/2022, returning all remaining knee and ankle UHMWPE devices labeled with an 8-year shelf life not packaged in EVOH/Nylon bags. A communication to healthcare professionals should follow. Instructions to recipients included to immediately identify and quarantine any affected product in inventory, notify customers down the line if further distributed, and complete and return the Recall Inventory Response Form to Exactech via email at recalls@exac.com. Exactech expanded their recall and sent an updated notification to US agents and distributors on 02/07/2022. On about 02/08/2022, a letter to surgeons and hospitals was sent. The notifications indicated that all non-conforming devices were included in the recall, to cease use of those devices, and work with sales agents to return any inventory. The letters to surgeons and hospitals also advised physicians regarding follow up with implanted patients.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026