Excelsior Medical Corp ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior Medical Corporation, Neptune, NJ. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior Medical Corporation, Neptune, NJ.
Brand
Excelsior Medical Corp
Lot Codes / Batch Numbers
All Serial Numbers since 1996.
Products Sold
All Serial Numbers since 1996.
Excelsior Medical Corp is recalling ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior due to Potential for alarm failure when the 140 pump completes infusion and/or the pump encounters a mid-infusion occlusion, when used in the 140 mode with a. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for alarm failure when the 140 pump completes infusion and/or the pump encounters a mid-infusion occlusion, when used in the 140 mode with a 140cc syringe. The ESP Pump works correctly when used in 60 mode with a 60cc syringe or smaller.
Recommended Action
Per FDA guidance
An 'Urgent Recall' letter was mailed certified return receipt mail on 3/22/06 to all customers per distribution list.