FHC, Inc. FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN)
Brand
FHC, Inc.
Lot Codes / Batch Numbers
All codes
Products Sold
All codes
FHC, Inc. is recalling FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power A due to Potential for non-sterile pin to contaminate sterile field. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential for non-sterile pin to contaminate sterile field
Recommended Action
Per FDA guidance
FHC notified customers by letter dated 3/14/05 and 3/25/05. The firm has issued a temporary set of instructions which include covering the non-sterile engagement pin with sterile tape and an amended Instruction For Use page to all customers. A redesigned pulley cover which will be sent to all customer when it becomes available in April 2005.