Flower Orthopedics Corporation Flower Drill Bit Kit 2.0 mm - intended to be used with solid 2.0mm screws and can be used with any plate that has been cleared as part of the Flower Small and Medium Implant Set and uses the small locking mechanism. mechanism. Catalog Number: DBK 020 2013002721 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Flower Drill Bit Kit 2.0 mm - intended to be used with solid 2.0mm screws and can be used with any plate that has been cleared as part of the Flower Small and Medium Implant Set and uses the small locking mechanism. mechanism. Catalog Number: DBK 020 2013002721
Brand
Flower Orthopedics Corporation
Lot Codes / Batch Numbers
Lot Number: 2013002721, 2013003589 Unique Identifier: 00840118101264
Products Sold
Lot Number: 2013002721, 2013003589 Unique Identifier: 00840118101264
Flower Orthopedics Corporation is recalling Flower Drill Bit Kit 2.0 mm - intended to be used with solid 2.0mm screws and can be used with any p due to Drill bits packaged in kits are missing the AO connect feature and may result in delay of procedure. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Drill bits packaged in kits are missing the AO connect feature and may result in delay of procedure
Recommended Action
Per FDA guidance
Coventus Orthopedic issued email communication on 9/01/21. Letter states reason for recall, health risk and action to take: Check your inventory for the product listed below. Next, please complete and return the attached Acknowledgement and Receipt Form by Friday, September 10, 2021. If you indicate on the form that you are in possession of affected product, Conventus Flower will ask that you return that affected kits. Conventus Flower will send replacement product as soon as we receive your completed Acknowledgement and Receipt Form. New DBK 020, DBK 024, DBK 027, or LSK 027 kits will be shipped to customers to replace affected product once the Acknowledgement and Receipt Form is received. Other Information: "Contact information for questions: Monday through Friday, 8:30 AM to 5:00 PM, EDST 877-778-8587 or FLOWERCS@FLOWERORTHO.COM