Digital Multimeters (Fluke) – Voltage Measurement (2002)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Consumer Product Safety Commission (CPSC)
Product
Fluke Digital Multimeters
Brand
Fluke Corp., of Everett, Wash.
Lot Codes / Batch Numbers
Not specified in CPSC notice. Check official source for details.
Product Images
Source: U.S. Consumer Product Safety Commission
Fluke Corp., of Everett, Wash. is recalling Fluke Digital Multimeters. The hazard is: The recalled units can take longer than normal, up to 18 seconds, to display readings of AC voltages. Based on CPSC recall notice.
Summary derived from CPSC notice
The following details were not provided by CPSC: UPC codes.
Reason for Recall
As stated by CPSC
Multimeters are used to measure voltage, resistance and current. The recalled Fluke digital multimeters measure up to 1,000 volts alternating current (VAC) and direct current (VDC). "Fluke" and "175", "177" or "179" are written on the front of the unit. The tester body is yellow and black, and measures about 7.25 inches long by 3.5 inches wide. Recalled units have a serial number below 79000000. The serial number is written on the back of the instrument, under the hinged stand.
Reported concern (CPSC)
The recalled units can take longer than normal, up to 18 seconds, to display readings of AC voltages above 500 volts. Users can misinterpret the delayed reading to mean that high voltage is not present. If high voltage is present, users could be exposed to a risk of shock, electrocution, and thermal burns.
Recommended Action
Per CPSC guidance
Consumers should stop using these testers immediately and contact Fluke for information on returning the recalled unit for a free repair.
Verify this information on the official source
For complete details and official instructions, check the U.S. Consumer Product Safety Commission (CPSC) directly.
View official CPSC recall noticePage updated: Jan 6, 2026