Fresenius Medical Care Holdings, Inc. Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343
Brand
Fresenius Medical Care Holdings, Inc.
Lot Codes / Batch Numbers
All Liberty Select Cyclers with Software Version 2.9.0 Model Number 180343 and RTLR180343
Products Sold
All Liberty Select Cyclers with Software Version 2.9.0 Model Number 180343 and RTLR180343
Fresenius Medical Care Holdings, Inc. is recalling Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343 due to The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the He. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may lead to inadequate dialysis which could result in the accumulation of toxins and fluid. Symptoms may include shortness of breath, increased weight gain, and increased blood pressure.
Recommended Action
Per FDA guidance
On September 16, 2020, Urgent Medical Device Correction notices and response forms were issued to home therapy nurse managers and patients. Patients using a Liberty Select cycler with software version 2.9.0 are instructed to disable the Flow Alert option. Patients will no longer receive on an-screen Flow Alert message or an audible Flow Alert series of beeps. Instructions for disabling the Flow Alert alarm are included in the notification letter. Patients should continue to use their Liberty Select Cycler for treatment with the Flow Alert option set to "No". Fresenius Medical Care LLC is not requesting the return of the Liberty Select cyclers software version 2.9.0. Replacing the Liberty Select cycler will not resolve this issue. Please complete and return the enclosed Reply Form, indicating receipt and understanding of the communication and actions to take. For additional information, please call 1-855-616-2309 Monday - Friday 8:30 am-5 pm, EST or submit an online request any time at www.fmcna-medinfo.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.