Fresenius Medical Care Renal Therapies Group, LLC Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as follows: (1) LIBERTY Line Cycler Set with Stay-safe connectors (Dual Patient Connector) Part Number: 050-87212 (2) LIBERTY Line Cycler Set with Stay-safe connectors (Single Patient Connector) Part Number: 050-87215 (3) LIBERTY Integrated Cycler Set with single Stay-safe patient connector and extended drain line Part Number: 050-87216 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as follows: (1) LIBERTY Line Cycler Set with Stay-safe connectors (Dual Patient Connector) Part Number: 050-87212 (2) LIBERTY Line Cycler Set with Stay-safe connectors (Single Patient Connector) Part Number: 050-87215 (3) LIBERTY Integrated Cycler Set with single Stay-safe patient connector and extended drain line Part Number: 050-87216
Brand
Fresenius Medical Care Renal Therapies Group, LLC
Lot Codes / Batch Numbers
All lots
Products Sold
All lots
Fresenius Medical Care Renal Therapies Group, LLC is recalling Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal di due to Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.
Recommended Action
Per FDA guidance
Fresenius issued on 05/29/19 The IMPORTANT MEDICAL DEVICE INFORMATION notification explains to the Home Therapy Nurses/Clinics to review the new flyer with each of their APD patients who use a Liberty Cycler. For the patient notification, it will encourage them to use and follow the attached instructional flyer during set up for their daily dialysis treatment. The notification will have a copy of the new step-by-step instructional flyer. Included is a Reply Form for confirmation the consignee read and understands the notice sent to them.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026