Galil Medical, Inc. Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip c Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip c
Brand
Galil Medical, Inc.
Lot Codes / Batch Numbers
System Serial Number / PM Kit #: VL0031 / PM0506M, VL0036 / PM0546M, VL0067 / PM0515M, VL0069 / PM0565M, VL0075 / PM0532M, VL0079 / PM0539M, VL0082 / PM0491M, VL0085 / PM0486M, VL0086 / PM0499M, VL0088 / PM0489M, VL0090 / PM0513M, VL0092 / PM0548M, VL0093 / PM0544M, VL0094 / PM0520M, VL0100 / PM0516M, VL0102 / PM0529M, VL0105 / PM0560M, VL0112 / PM0542M, VL0115 / PM0525M, VL0136 / PM0561M, VL0148 / PM0514M, VL0153 / PM0531M, VL0171 / PM0564M, VL0175 / PM0559M, VL0181 / PM0512M, VL0186 / PM0533M, VL0191 / PM0492M, VL0220 / PM0545M, VL0230 / PM0528M, VL0238 / PM0500M, VL0249 / PM0498M, VL0271 / PM0497M, VL0275 / PM0543M, VL0280 / PM0549M, VL0283 / PM0488M, VL0285 / PM0566M, VL0287 / PM0562M, VL0291 / PM0524M, VL0300 / PM0550M, VL0301 / PM0541M, VL0303 / PM0530M, VL0304 / PM0563M, VL0308 / PM0523M, VL0309 / PM0507M, VL0310 / PM0490M, VL0311 / PM0555M, VL0312 / PM0526M, VL0321 / PM0552M, VL0362 / PM0527M.
Products Sold
System Serial Number / PM Kit #: VL0031 / PM0506M, VL0036 / PM0546M, VL0067 / PM0515M, VL0069 / PM0565M, VL0075 / PM0532M, VL0079 / PM0539M, VL0082 / PM0491M, VL0085 / PM0486M, VL0086 / PM0499M, VL0088 / PM0489M, VL0090 / PM0513M, VL0092 / PM0548M, VL0093 / PM0544M, VL0094 / PM0520M, VL0100 / PM0516M, VL0102 / PM0529M, VL0105 / PM0560M, VL0112 / PM0542M, VL0115 / PM0525M, VL0136 / PM0561M, VL0148 / PM0514M, VL0153 / PM0531M, VL0171 / PM0564M, VL0175 / PM0559M, VL0181 / PM0512M, VL0186 / PM0533M, VL0191 / PM0492M, VL0220 / PM0545M, VL0230 / PM0528M, VL0238 / PM0500M, VL0249 / PM0498M, VL0271 / PM0497M, VL0275 / PM0543M, VL0280 / PM0549M, VL0283 / PM0488M, VL0285 / PM0566M, VL0287 / PM0562M, VL0291 / PM0524M, VL0300 / PM0550M, VL0301 / PM0541M, VL0303 / PM0530M, VL0304 / PM0563M, VL0308 / PM0523M, VL0309 / PM0507M, VL0310 / PM0490M, VL0311 / PM0555M, VL0312 / PM0526M, VL0321 / PM0552M, VL0362 / PM0527M.
Galil Medical, Inc. is recalling Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system inten due to Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing pe. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance.
Recommended Action
Per FDA guidance
An Urgent Medical Device Correction letter was emailed to customers beginning 10/25/2019. A hard copy of this letter was sent by USPS. The letter identified the affected devices, stated the issue, instructed not to use the IceSeed 1.5 or Ice Rod 1.5 PLUS needles until a preventive maintenance activity, to replace the desiccant tubes, has been done by an authorized service engineer.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AK, AZ, AR, CA, FL, GA, HI, KY, MA, MI, MN, NY, NC, ND, OH, PA, SC, TN, UT, VA, PR