GE Healthcare GE Healthcare Innova 3100 Digital Fluoroscopic Imaging System with Bolus Chasing option. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
GE Healthcare Innova 3100 Digital Fluoroscopic Imaging System with Bolus Chasing option.
Brand
GE Healthcare
Lot Codes / Batch Numbers
8613100, 77573841, 0910064009, 3018774100, 4103323100, 4103374100, 4103623100, 4105434100, 4438494100, 5138534100, 5703214100, 7083333100, 7278253100, 8019643100, 8595724100, 9408984100, 082416040017, 082416070004, 082416120006, 082416130013, 082416140003, 201541LAB2, 229228INNOV41V, 239436CATH1, 251435ICV5, 270688INN, 304255INNOVA, 304347INNOVA, 309691P3100, 310423INNOVA1, 313593IC1, 313593IC2, 314525SACL2, 315448RM4, 315448RM5, 316268INNOVAA, 319338VA4100, 325670CATH, 334286BSP, 352333VASC, 352376SP, 352401CATH, 352401VASC, 378INV31, 407SPECIALS, 409899LAB3, 413447CATH, 414961INN4100, 416756INNOVA, 417269SCLD, 478633INOV, 502587LAB4, 508856OR41, 516562INNOVA4, 516663INNOVA2, 518525LAB1, 540953INNOVA, 561266IN31, 563421CATHLAB8, 573632CRA, 573761SMA, 573815BCR1, 574535IN41, 574647IN41, 6012884100SP1, 601553CL3, 602239LAB4, 602239LAB6, 603778ECL1, 606833INNOVA, 614293LIBERTY1, 614293LIBERTY2, 615338CV2, 615769SKYCV2, 619340VA8, 619482INNOVA1, 619541INNOVA31, 619552INNOVA1, 626359INNOVA, 702616INNOVA, 7028803100A, 706774CCL5, 713794LAB7, 713794LAB9, 715369INN4100, 717217INN31, 717763INNOVA41, 717763OR4100, 718283INNOVA3, 718518CVL1, 724656ANG, 724773C1, 727725SPEC1, 731422CV1, 732294INOV, 757398INNOVA, 757736INNOVA, 765453INNOVA1, 770991CL3, 773792CL3100, 773NORWG3100, 803329CL31, 803329CL41, 804594CATH, 808547QCV2A, 812238LIB1, 812279INNOVA1, 812283IN31, 812450EP1, 812482INNOVA, 813972INTV, 816276SP1, 817255NH4100, 817433HSW3100, 83INNOVA1188379, 845365V19, 850678IN3100, 865541CCL5, 865693CCL3, 901726SP2, 903CSM4100, 904202CATH3, 904276SP, 908522LAB1, 909788INNOVA1, 918307SHIN4, 918744JC4100, 928779INNOVA1, 936266CL1, 956389INN31, 956632INN4100, 956994INN1, 956994INN2, 972487ANGIO1, 973971LAB3, BPINNOVA2, BPINNOVA3R, BPINNOVA3R, ECXV1406C, FHORLCL1, HC1536XR04, RU1550VA01, RU2457VA01, SLIN4100, YV0022, and YV0036.
Products Sold
8613100, 77573841, 0910064009, 3018774100, 4103323100, 4103374100, 4103623100, 4105434100, 4438494100, 5138534100, 5703214100, 7083333100, 7278253100, 8019643100, 8595724100, 9408984100, 082416040017, 082416070004, 082416120006, 082416130013, 082416140003, 201541LAB2, 229228INNOV41V, 239436CATH1, 251435ICV5, 270688INN, 304255INNOVA, 304347INNOVA, 309691P3100, 310423INNOVA1, 313593IC1, 313593IC2, 314525SACL2, 315448RM4, 315448RM5, 316268INNOVAA, 319338VA4100, 325670CATH, 334286BSP, 352333VASC, 352376SP, 352401CATH, 352401VASC, 378INV31, 407SPECIALS, 409899LAB3, 413447CATH, 414961INN4100, 416756INNOVA, 417269SCLD, 478633INOV, 502587LAB4, 508856OR41, 516562INNOVA4, 516663INNOVA2, 518525LAB1, 540953INNOVA, 561266IN31, 563421CATHLAB8, 573632CRA, 573761SMA, 573815BCR1, 574535IN41, 574647IN41, 6012884100SP1, 601553CL3, 602239LAB4, 602239LAB6, 603778ECL1, 606833INNOVA, 614293LIBERTY1, 614293LIBERTY2, 615338CV2, 615769SKYCV2, 619340VA8, 619482INNOVA1, 619541INNOVA31, 619552INNOVA1, 626359INNOVA, 702616INNOVA, 7028803100A, 706774CCL5, 713794LAB7, 713794LAB9, 715369INN4100, 717217INN31, 717763INNOVA41, 717763OR4100, 718283INNOVA3, 718518CVL1, 724656ANG, 724773C1, 727725SPEC1, 731422CV1, 732294INOV, 757398INNOVA, 757736INNOVA, 765453INNOVA1, 770991CL3, 773792CL3100, 773NORWG3100, 803329CL31, 803329CL41, 804594CATH, 808547QCV2A, 812238LIB1, 812279INNOVA1, 812283IN31, 812450EP1, 812482INNOVA, 813972INTV, 816276SP1, 817255NH4100, 817433HSW3100, 83INNOVA1188379, 845365V19, 850678IN3100, 865541CCL5, 865693CCL3, 901726SP2, 903CSM4100, 904202CATH3, 904276SP, 908522LAB1, 909788INNOVA1, 918307SHIN4, 918744JC4100, 928779INNOVA1, 936266CL1, 956389INN31, 956632INN4100, 956994INN1, 956994INN2, 972487ANGIO1, 973971LAB3, BPINNOVA2, BPINNOVA3R, BPINNOVA3R, ECXV1406C, FHORLCL1, HC1536XR04, RU1550VA01, RU2457VA01, SLIN4100, YV0022, and YV0036.
GE Healthcare is recalling GE Healthcare Innova 3100 Digital Fluoroscopic Imaging System with Bolus Chasing option. due to Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c
Recommended Action
Per FDA guidance
GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.