GE Healthcare GE LightSpeed VCT Select / LightSpeed Pro32, Model 5126093, CT Scanner system. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
GE LightSpeed VCT Select / LightSpeed Pro32, Model 5126093, CT Scanner system.
Brand
GE Healthcare
Lot Codes / Batch Numbers
All serial numbers
Products Sold
All serial numbers
GE Healthcare is recalling GE LightSpeed VCT Select / LightSpeed Pro32, Model 5126093, CT Scanner system. due to Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.
Recommended Action
Per FDA guidance
A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, DC, PR