GE Healthcare Integrated IT Solutions GE Centricity PACS (Picture Archiving and Communication System) software; GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity PACS is used to receive, store, distribute, display, manipulate, edit and annotate images throughout a clinical environment. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
GE Centricity PACS (Picture Archiving and Communication System) software; GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity PACS is used to receive, store, distribute, display, manipulate, edit and annotate images throughout a clinical environment.
Brand
GE Healthcare Integrated IT Solutions
Lot Codes / Batch Numbers
Centricity PACS software versions 2.1.X and 3.0.X and Centricity PACS versions 1.0.X and 2.0.X.
Products Sold
Centricity PACS software versions 2.1.X and 3.0.X and Centricity PACS versions 1.0.X and 2.0.X.
GE Healthcare Integrated IT Solutions is recalling GE Centricity PACS (Picture Archiving and Communication System) software; GE Healthcare Integrated I due to There is a potential safety issue associated with the use of GE Centricity PACS software related to configuration of DICOM query retrieval with reject. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential safety issue associated with the use of GE Centricity PACS software related to configuration of DICOM query retrieval with rejected images.
Recommended Action
Per FDA guidance
GE Healthcare Integrated IT Solutions sent an "Urgent Medical Device Correction" letter dated April 2, 2009 to their Centricity PACS software versions 2.1.x and 3.0.x customers with attention to the Director of Radiology, the Hospital Administrator, the Head of the Radiology Department and the PACS Administrator. The letters advised users of the patient safety issue associated with the affected device and provided safety instructions to follow until the software is updated. User were asked to complete the enclosed Customer Reply form and fax the completed form to 1-847-939-1479. Centricity PACS Software Versions 1.0.x and 2.0.x customers were sent separate letters explaining the issue and the safety instructions to follow, but there will be no software updates since those software versions are no longer being supported. Any questions should be directed to the Remote Online Center, Field Engineer or the GE Customer Care Center at 1-800-437-1171.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026