GE Healthcare, LLC Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo regulate based on their own physiology. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo regulate based on their own physiology.
Brand
GE Healthcare, LLC
Lot Codes / Batch Numbers
Model Number: 2082844-001, Serial Number (UDI Number): TABY70690 (010084068211686221TABY7069011190610), TABY70692 (010084068211686221TABY7069211190611), TABY70708 (010084068211686221TABY7070811190612), TABY70719 (010084068211686221TABY7071911190613), TABY70709 (010084068211686221TABY7070911190612), TABY70724 (010084068211686221TABY7072411190614), TABY70722 (010084068211686221TABY7072211190614), TABY70718 (010084068211686221TABY7064211190603), TABY70717 (010084068211686221TABY7064411190603), TABY70712 (010084068211686221TABY7064511190603), TABY70678 (010084068211686221TABY7067811190607), TABY70725 (010084068211686221TABY7072511190615), TABY70672 (010084068211686221TABY7067211190607), TABY70686 (010084068211686221TABY7068611190610), TABY70688 (010084068211686221TABY7068811190610), TABY70697 (010084068211686221TABY7069711190611), TABY70726 (010084068211686221TABY7072611190616), TABY70729 (010084068211686221TABY7072911190616), TABY70645 (010084068211686221TABY7064711190604), TABY70647 (010084068211686221TABY7066011190605), TABY70646 (010084068211686221TABY7071211190613), TABY70644 (010084068211686221TABY7071711190613), TABY70642 (010084068211686221TABY7071811190613), TABY70732 (010084068211686221TABY7073211190617), TABY70723 (010084068211686221TABY7072311190614), TABY70747 (010084068211686221TABY7074711190619), TABY70748 (010084068211686221TABY7074811190619), TABY70673 (010084068211686221TABY7067311190607), TABY70676 (010084068211686221TABY7067611190607), TABY70693 (010084068211686221TABY7069311190611), TABY70648 (010084068211686221TABY7064811190604), TABY70675 (010084068211686221TABY7067511190607), TABY70694 (010084068211686221TABY7069411190611), TABY70695 (010084068211686221TABY7069511190611), TABY70730 (010084068211686221TABY7073011190617), TABY70734 (010084068211686221TABY7073411190617), TABY70737 (010084068211686221TABY7073711190618), TABY70677 (010084068211686221TABY7067711190607), TABY70698 (010084068211686221TABY7069811190611), TABY70699 (010084068211686221TABY7069911190611), TABY70703 (010084068211686221TABY7070311190612), TABY70706 (010084068211686221TABY7070611190612), TABY70727 (010084068211686221TABY7072711190616), TABY70728 (010084068211686221TABY7072811190616), TABY70660 (010084068211686221TABY7064611190603), TABY70689 (010084068211686221TABY7068911190610), TABY70705 (010084068211686221TABY7070511190612), TABY70707 (010084068211686221TABY7070711190612), TABY70731 (010084068211686221TABY7073111190617), TABY70733 (010084068211686221TABY7073311190617), TABY70652 (Not Available), TABY70654 (Not Available), TABY70662 (Not Available), TABY70664 (Not Available), TABY70710 (Not Available), TABY70715 (Not Available), TABY70716 (Not Available), TABY70753 (Not Available), TABY70659 (Not Available), TABY70704 (010084068211686221TABY7070411190612), TABY70755 (010084068211686221TABY7075511190620), TABY70638 (Not Available), TABY70649 (Not Available), TABY70656 (Not Available), TABY70661 (Not Available), TABY70657 (Not Available), TABY70650 (Not Available), TABY70667 (Not Available), TABY70668 (Not Available), TABY70653 (Not Available), TABY70663 (Not Available), TABY70670 (Not Available), TABY70665 (Not Available), TABY70669 (Not Available), TABY70671 (Not Available), TABY70714 (Not Available), TABY70658 (Not Available), TABY70655 (Not Available), TABY70666 (Not Available), TABY70651 (Not Available).
Products Sold
Model Number: 2082844-001; Serial Number (UDI Number): TABY70690 (010084068211686221TABY7069011190610), TABY70692 (010084068211686221TABY7069211190611), TABY70708 (010084068211686221TABY7070811190612), TABY70719 (010084068211686221TABY7071911190613), TABY70709 (010084068211686221TABY7070911190612), TABY70724 (010084068211686221TABY7072411190614), TABY70722 (010084068211686221TABY7072211190614), TABY70718 (010084068211686221TABY7064211190603), TABY70717 (010084068211686221TABY7064411190603), TABY70712 (010084068211686221TABY7064511190603), TABY70678 (010084068211686221TABY7067811190607), TABY70725 (010084068211686221TABY7072511190615), TABY70672 (010084068211686221TABY7067211190607), TABY70686 (010084068211686221TABY7068611190610), TABY70688 (010084068211686221TABY7068811190610), TABY70697 (010084068211686221TABY7069711190611), TABY70726 (010084068211686221TABY7072611190616), TABY70729 (010084068211686221TABY7072911190616), TABY70645 (010084068211686221TABY7064711190604), TABY70647 (010084068211686221TABY7066011190605), TABY70646 (010084068211686221TABY7071211190613), TABY70644 (010084068211686221TABY7071711190613), TABY70642 (010084068211686221TABY7071811190613), TABY70732 (010084068211686221TABY7073211190617), TABY70723 (010084068211686221TABY7072311190614), TABY70747 (010084068211686221TABY7074711190619), TABY70748 (010084068211686221TABY7074811190619), TABY70673 (010084068211686221TABY7067311190607), TABY70676 (010084068211686221TABY7067611190607), TABY70693 (010084068211686221TABY7069311190611), TABY70648 (010084068211686221TABY7064811190604), TABY70675 (010084068211686221TABY7067511190607), TABY70694 (010084068211686221TABY7069411190611), TABY70695 (010084068211686221TABY7069511190611), TABY70730 (010084068211686221TABY7073011190617), TABY70734 (010084068211686221TABY7073411190617), TABY70737 (010084068211686221TABY7073711190618), TABY70677 (010084068211686221TABY7067711190607), TABY70698 (010084068211686221TABY7069811190611), TABY70699 (010084068211686221TABY7069911190611), TABY70703 (010084068211686221TABY7070311190612), TABY70706 (010084068211686221TABY7070611190612), TABY70727 (010084068211686221TABY7072711190616), TABY70728 (010084068211686221TABY7072811190616), TABY70660 (010084068211686221TABY7064611190603), TABY70689 (010084068211686221TABY7068911190610), TABY70705 (010084068211686221TABY7070511190612), TABY70707 (010084068211686221TABY7070711190612), TABY70731 (010084068211686221TABY7073111190617), TABY70733 (010084068211686221TABY7073311190617), TABY70652 (Not Available), TABY70654 (Not Available), TABY70662 (Not Available), TABY70664 (Not Available), TABY70710 (Not Available), TABY70715 (Not Available), TABY70716 (Not Available), TABY70753 (Not Available), TABY70659 (Not Available), TABY70704 (010084068211686221TABY7070411190612), TABY70755 (010084068211686221TABY7075511190620), TABY70638 (Not Available), TABY70649 (Not Available), TABY70656 (Not Available), TABY70661 (Not Available), TABY70657 (Not Available), TABY70650 (Not Available), TABY70667 (Not Available), TABY70668 (Not Available), TABY70653 (Not Available), TABY70663 (Not Available), TABY70670 (Not Available), TABY70665 (Not Available), TABY70669 (Not Available), TABY70671 (Not Available), TABY70714 (Not Available), TABY70658 (Not Available), TABY70655 (Not Available), TABY70666 (Not Available), TABY70651 (Not Available).
GE Healthcare, LLC is recalling Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a due to Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.
Recommended Action
Per FDA guidance
GE Healthcare notified customers on about 07/26/2019 via phone calls and a "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were instructed to immediately stop use of and segregate the affected devices, that GE Healthcare would correct all affected products at no cost, and a GE Healthcare representative will contact them to arrange for the correction. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative. A second customer letter was sent on 08/05/2019 to all customer. The second letter contained the same instructions, but provided an updated list of affected serial numbers.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026