GE Healthcare, LLC Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.
Brand
GE Healthcare, LLC
Lot Codes / Batch Numbers
Serial numbers: GSHW60086, GSHX60001, GSHX60014, GSHY70002-GSHY70195, GSHZ70101-GSHZ70108, MAAP50034, MAAQ50170, MAAR50021, MAAT50013, MAAT60045, MAAU61029, MAAU61040 and MAAV6059.
Products Sold
Serial numbers: GSHW60086, GSHX60001, GSHX60014, GSHY70002-GSHY70195, GSHZ70101-GSHZ70108, MAAP50034, MAAQ50170, MAAR50021, MAAT50013, MAAT60045, MAAU61029, MAAU61040 and MAAV6059.
GE Healthcare, LLC is recalling Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808. due to During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Recommended Action
Per FDA guidance
GE Healthcare notified customers on 04/02/2020 via "Urgent Medical Device Correction" letter. The recall letter identified the affected product and serial numbers. The firm instructed the users to discontinue use of the product if at any time they smell a gas odor. GE Healthcare will correct all affected products at a later date when the correction is available at no cost.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.