GE Healthcare, LLC Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Brand
GE Healthcare, LLC
Lot Codes / Batch Numbers
Serial Numbers: 64GY51017, 64GY51020, 64GY51022, 64GZ51028, 64GZ51030, 64GZ51029, 64GY51021
Products Sold
Serial Numbers: 64GY51017, 64GY51020, 64GY51022, 64GZ51028, 64GZ51030, 64GZ51029, 64GY51021
GE Healthcare, LLC is recalling Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging due to GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four roto. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Recommended Action
Per FDA guidance
The firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.