GE Healthcare The LightSpeed RT16 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. The system is capable of generating images for the guidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology. The system allows imaging of obese patients, up to and including the obese population (BMI > 40). When used in the LightSpeed RT 16 configuration, the system acquires CT anatomical images that are clinically useful in the simulation and Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
The LightSpeed RT16 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. The system is capable of generating images for the guidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology. The system allows imaging of obese patients, up to and including the obese population (BMI > 40). When used in the LightSpeed RT 16 configuration, the system acquires CT anatomical images that are clinically useful in the simulation and
Brand
GE Healthcare
Lot Codes / Batch Numbers
All serial numbers
Products Sold
All serial numbers
GE Healthcare is recalling The LightSpeed RT16 CT Scanner System is indicated for head and whole body X-ray Computed Tomography due to When the manual cradle release feature on the table is used to move a patient out of the gantry, a discrepancy of up to 6mm between the numerical disp. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When the manual cradle release feature on the table is used to move a patient out of the gantry, a discrepancy of up to 6mm between the numerical display on the gantry and the actual position of the table could be introduced.
Recommended Action
Per FDA guidance
A Product Safety Advisory Notice, dated December 8, 2006, was issued to all customers informing them of the safety issue and production correction. Customers are advised to operate the CT cradle as intended with the system''s electronic controls. If a critical need arises to release the cradle during a radiation therapy simulation scan, they are advised to independently verify patient position prior to marking the patient for therapy to avoid the potential inaccuracy. A software update will be installed to correct this problem by a GE representative.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution