LOGIQ P9 Ultrasound (GE Medical) – Measurement Inaccuracy (2025)
Severity assessment pending.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
LOGIQ P9 R4.5 ultrasound system with system software version R4.5.7., Model Number 5877533
Brand
GE Medical Systems, LLC
Lot Codes / Batch Numbers
Model 5877533 UDI-DI: 00195278464637
Products Sold
Model 5877533 UDI-DI: 00195278464637, Serial Numbers: LP9490416 LP9490175 LP9490393 LP9490399 LP9490376 LP9490463 LP9443017 LP9442472 LP9443231 LP9442094 LP9443291 LP9443230 LP9440403 LP9442483 LP9440972 LP9440166 LP9440237 LP9440933 LP9440348 LP9442196 LP9440631 LP9440926 LP9441099 LP9441225 LP9441097 LP9440929 LP9441030 LP9440921 LP9441035 LP9440507 LP9440920 LP9440459 LP9440460 LP9441332 LP9442656 LP9443022 LP9440986
GE Medical Systems, LLC is recalling LOGIQ P9 R4.5 ultrasound system with system software version R4.5.7., Model Number 5877533 due to The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
Recommended Action
Per FDA guidance
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers. Actions to be taken by Customer/User For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features. After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.74084@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Verify this information on the official source
This page aggregates publicly available data. Always confirm recall details directly with the U.S. Food and Drug Administration (FDA) before taking action.
View official FDA recall noticePage updated: Jan 10, 2026
Important Notice
This page displays information sourced from the U.S. Food and Drug Administration (FDA). IsMyProductRecalled.com aggregates public data for convenience and does not issue recalls. We are not affiliated with any government agency. Always verify recall information with the official source before taking action. This is not medical or legal advice.