General Electric Med Systems LLC Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2)
Brand
General Electric Med Systems LLC
Lot Codes / Batch Numbers
100943VP0, 67392II1, 36521VE3, 70827VE1, 419453BU6436332BU1, 70800VE8, 44586VE6, 341GE8, 97197VP8, 99813VP8, 97983VP1, 70820VE6, 58626VE3, 70795VE0, FRB3510J25, 58630VE5, 58624VE8, 58589VE3, 7823VE0, 44579VE1, 44597VE3, 44585VE8, 25125VE6, 425022BU1, 70796VE8, 93086VE7, 37621BU4, 76906VE7, 76899VE4, 70798VE4, 65504VE3, 65603VE3, 36543VE7, 58628VE9, 70811VE5, 44584VE1, 44581VE7, FRB3430VBM, 44589VE0, 44598VE1, 44582VE5, 44594VE0, 72740VE4, 1189GE0, 2030GE5, 2085GE0, 2025GE5, 72477II3, 72480II7, 343GE4, 97203VP4, 98272VP8, 423277BU3, 349GE1, 425018BU9, FRB3300M6G, 99699VP1, 98371VP8, 101216VP0, 425015BU5, 237GE8, 364GE0, 339GE2, 101196VP4, 419382BU7, 419383BU5, 366GE5, 367GE3, 357GE4, 2085GE9, 419457BU7, 25128VE0, 98436VP9, 99981VP3, 36513VE0, 98623UP2, 101000VP8, 2028GE9, 2799GE5, 1089GE2, 419450BU2, 101280VP6, 67269VE1, 100855VP6, 66250II2, 346GE7, 363GE2, 345GE9, 1073VP5, 101170VP9, 98516VP8, 2802GE7, 419356BU1, 66244II5, 1487GE8, 67389II7, 365GE7, 425021BU3, 383GE0, 25132VE2, 101334VP1, 95972VP6, 419449BU4, 2031GE3, 2800GE1, 36517VE1, FRB33605L9, 1024GE9, 95307VP5, 102235VP9, 102232VP6, 425204BU5, 1124GE7, 1122GE1, 419455BU1, 419456BU9, 102220VP1, 102555VP0, 97622VP5, 72249II6, 67210VE5, FRB33605KX, 419257BU1, 95268VP9, 338GE4, 101404VP2, 2090GE9, 2087GE5, 101308VP5, 101251VP7, 421989BU5, 342GE6, 99627VP2, 97933VP6, 98653VP9, 66249II4, 72479II9, 25126VE4, 356GE6, 25130VE6, 65450VE9, DE31500056, 347GE5, 421367BU4, 100912VP5, 25123VE1, 25124VE9, 99867VP4, DE24600031, 99910VP2, 101246VP7, 419357BU9, 101171VP7, 102825VP7, 100664VP2, 102856VP2, 424101BU4, 425020BU5, 354GE1, 36518VE9, 100885VP3, 2801GE9, USV41104ST, 2091GE7, USV41103MJ, 99994VP6, 102133VP6, 419549BU1, 25137VE1, 102078VP3, 425351BU4, 1066GE0, 98816VP2, 97591VP2, 99631VP4, 423318BU5, 2086GE7, 97905VP4, 100690VP7, 350GE9, 25139VE7, 1120GE5, 100001VP7, 2084GE2, 391228BU4, 101190VP7, 99909VP4, 1123GE9, 2803GE5, 98474VP0, 25127VE2, 67384II8, 98815VP4, 101288VP9, 101287VP1, 66241II1, 25134VE8
Products Sold
100943VP0, 67392II1, 36521VE3, 70827VE1, 419453BU6436332BU1, 70800VE8, 44586VE6, 341GE8, 97197VP8, 99813VP8, 97983VP1, 70820VE6, 58626VE3, 70795VE0, FRB3510J25, 58630VE5, 58624VE8, 58589VE3, 7823VE0, 44579VE1, 44597VE3, 44585VE8, 25125VE6, 425022BU1, 70796VE8, 93086VE7, 37621BU4, 76906VE7, 76899VE4, 70798VE4, 65504VE3, 65603VE3, 36543VE7, 58628VE9, 70811VE5, 44584VE1, 44581VE7, FRB3430VBM, 44589VE0, 44598VE1, 44582VE5, 44594VE0, 72740VE4, 1189GE0, 2030GE5, 2085GE0, 2025GE5, 72477II3, 72480II7, 343GE4, 97203VP4, 98272VP8, 423277BU3, 349GE1, 425018BU9, FRB3300M6G, 99699VP1, 98371VP8, 101216VP0, 425015BU5, 237GE8, 364GE0, 339GE2, 101196VP4, 419382BU7, 419383BU5, 366GE5, 367GE3, 357GE4, 2085GE9, 419457BU7, 25128VE0, 98436VP9, 99981VP3, 36513VE0, 98623UP2, 101000VP8, 2028GE9, 2799GE5, 1089GE2, 419450BU2, 101280VP6, 67269VE1, 100855VP6, 66250II2, 346GE7, 363GE2, 345GE9, 1073VP5, 101170VP9, 98516VP8, 2802GE7, 419356BU1, 66244II5, 1487GE8, 67389II7, 365GE7, 425021BU3, 383GE0, 25132VE2, 101334VP1, 95972VP6, 419449BU4, 2031GE3, 2800GE1, 36517VE1, FRB33605L9, 1024GE9, 95307VP5, 102235VP9, 102232VP6, 425204BU5, 1124GE7, 1122GE1, 419455BU1, 419456BU9, 102220VP1, 102555VP0, 97622VP5, 72249II6, 67210VE5, FRB33605KX, 419257BU1, 95268VP9, 338GE4, 101404VP2, 2090GE9, 2087GE5, 101308VP5, 101251VP7, 421989BU5, 342GE6, 99627VP2, 97933VP6, 98653VP9, 66249II4, 72479II9, 25126VE4, 356GE6, 25130VE6, 65450VE9, DE31500056, 347GE5, 421367BU4, 100912VP5, 25123VE1, 25124VE9, 99867VP4, DE24600031, 99910VP2, 101246VP7, 419357BU9, 101171VP7, 102825VP7, 100664VP2, 102856VP2, 424101BU4, 425020BU5, 354GE1, 36518VE9, 100885VP3, 2801GE9, USV41104ST, 2091GE7, USV41103MJ, 99994VP6, 102133VP6, 419549BU1, 25137VE1, 102078VP3, 425351BU4, 1066GE0, 98816VP2, 97591VP2, 99631VP4, 423318BU5, 2086GE7, 97905VP4, 100690VP7, 350GE9, 25139VE7, 1120GE5, 100001VP7, 2084GE2, 391228BU4, 101190VP7, 99909VP4, 1123GE9, 2803GE5, 98474VP0, 25127VE2, 67384II8, 98815VP4, 101288VP9, 101287VP1, 66241II1, 25134VE8
General Electric Med Systems LLC is recalling Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator due to Display in fused mode- viewer allows 2 studies ( PET/CT) registered in the same spatial domain to be viewed simultaneously. These 2 images should be . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Display in fused mode- viewer allows 2 studies ( PET/CT) registered in the same spatial domain to be viewed simultaneously. These 2 images should be visually aligned to a precision of less than 1/2 PET Voxel. A problem associated with this version of software may introduce a shift of up to 1.5 PET Voxels, which can be as much as 8 mm with certain zoom factors.
Recommended Action
Per FDA guidance
An Urgent Safety Notice letter, dated October 8, 2004,(5121309-147, rev.E) was distributed to affected customers and a validated software patch that corrected the problem was installed on affected systems via Field Modification Instruction (FMI) 80114.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AK, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MD, MA, MI, MN, MO, NH, NJ, NY, NC, ND, OH, OK, PA, SC, TN, TX, UT, VA, WA, WV, WI
Page updated: Jan 10, 2026