Getinge/Castle, Inc. Model 8666 Washer/Disinfector with holding tank heater option. Firm on label is Getinge Disinfection AB, Vaxjo, Sweden. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Model 8666 Washer/Disinfector with holding tank heater option. Firm on label is Getinge Disinfection AB, Vaxjo, Sweden.
Brand
Getinge/Castle, Inc.
Lot Codes / Batch Numbers
Serial numbers 81972 through 82063, and SEV0141098 through SEV0304112.
Products Sold
Serial numbers 81972 through 82063; and SEV0141098 through SEV0304112.
Getinge/Castle, Inc. is recalling Model 8666 Washer/Disinfector with holding tank heater option. Firm on label is Getinge Disinfecti due to Hot water may splash from the washer chamber creating a burn hazard.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Hot water may splash from the washer chamber creating a burn hazard.
Recommended Action
Per FDA guidance
Letter dated 3/25/03 flagged: ''Important Safety Alert/Product Upgrade''. Customers advised to discontinue use until the field correction is performed.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026