Getinge Usa Sales Inc Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.
Brand
Getinge Usa Sales Inc
Lot Codes / Batch Numbers
Part Number 6682282, UDI/DI: 07325710000212: Serial Numbers: 17003 - 23454.
Products Sold
Part Number 6682282; UDI/DI: 07325710000212: Serial Numbers: 17003 - 23454.
Getinge Usa Sales Inc is recalling Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, F due to Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalati. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Recommended Action
Per FDA guidance
On May 1, 2024 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken by customer: Until further notice, do NOT use the Vaproizer with Sevoflurane provided by Poramal and Baxter Sevoflurane. Do NOT use the Vaporizer if it was ever used with Sevoflurane Piramal and/or Baxter Sevoflurane, even if it currently is used with AbbVie Ultane. Our records indicate that you have received the Vaporizer Sevoflurane, Maquet Filling having one or more of the serial numbers that are affected by this recall. 1. Please examine your inventory immediately to determine if you have any of the Vaporizer Sevoflurane, Maquet Filling with the product codes/serial numbers listed in this notice. The Product Code/REF Number and serial number (SN) are found on the label located underneath each vaporizer. Should you have any affected product, please remove from areas of use. 2. If you have ONLY used Vaporizer Product Code/REF Number 6682282, serial number 17003-23454 with AbbVie Ultane, you can continue to use the vaporizer, by certifying that you only use AbbVie Ultane on the Medical Device Recall Response Form. If you are uncertain whether you have previously used the vaporizer with Piramal Sevoflurane and/or Baxter Sevoflurane, discontinue use of the vaporizer. 3. Persuant to the previous recall for the Vaporizer Product Code/REF Number 6886601, you can continue to use this vaporizer so long as it has ONLY been used with Abbvie Ultane. If you are uncertain whether you have previously used the vaporizer with Piramal Sevoflurane and/or Baxter Sevoflurane, discontinue use of the vaporizer. 4. If at any time, the agent in the vaporizer looks cloudy and yellowish, you should immediately stop use of the affected vaporizer. 5. If you have affected product you are entitled to a credit or replacement. Please contact Getinge Customer Service at 888-943-8872 (press option 2) to request a return authorization (RMA) and shipping instructions to return any affected product. Pack
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, CO, IN, MO, OR, VA, WA