GN Otometrics Horskatten 9 Taastrup Denmark Gold Cup Leadwire and Snap Leadwire with the following component descriptions: 1. 24" Snap Leadwire, Blue; 2. 40" Snap Leadwire, Blue; 3. 24" Snap Leadwire, Red; 4. 40" Snap Leadwire, Red; 5. 24" Snap Leadwire, Green; 6. 40" Snap Leadwire, Green; 7. 24" Snap Leadwire, Brown; 8. 40" Snap Leadwire, Brown; 9. 24" Snap Leadwire, Orange; 10. 40" Snap Leadwire, Orange; 11. 24" Snap Leadwire, White; 12. 40" Snap Leadwire, White; 13. 24" Sna Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Gold Cup Leadwire and Snap Leadwire with the following component descriptions: 1. 24" Snap Leadwire, Blue; 2. 40" Snap Leadwire, Blue; 3. 24" Snap Leadwire, Red; 4. 40" Snap Leadwire, Red; 5. 24" Snap Leadwire, Green; 6. 40" Snap Leadwire, Green; 7. 24" Snap Leadwire, Brown; 8. 40" Snap Leadwire, Brown; 9. 24" Snap Leadwire, Orange; 10. 40" Snap Leadwire, Orange; 11. 24" Snap Leadwire, White; 12. 40" Snap Leadwire, White; 13. 24" Sna
Brand
GN Otometrics Horskatten 9 Taastrup Denmark
Lot Codes / Batch Numbers
Part Numbers: 1. 7590310-24, 2. 7590310-40, 3. 7590311-24, 4. 7590311-40, 5. 7590312-24, 6. 7590312-40, 7. 7590313-24, 8. 7590313-40, 9. 7590314-24, 10. 7590314-40, 11. 7590315-24, 12. 7590315-40, 13. 7590316-24, 14. 7590316-40, 15. 7590317-24, 16. 7590317-40, 17. 7590318-24-5, 18. 7590318-24-7, 19. 7590318-40-5, 20. 7590318-40-7, 21. 7590319-24-5, 22. 7590319-40-5, 23. 7590440, 24. 7590441, 25. 7590442, 26. 7590443, 27. 7590444, 28. 7590445, 29. 7590446, 30. 7590447, 31. 7590450-5, 32. 7590450-8
Products Sold
Part Numbers: 1. 7590310-24; 2. 7590310-40; 3. 7590311-24; 4. 7590311-40; 5. 7590312-24; 6. 7590312-40; 7. 7590313-24; 8. 7590313-40; 9. 7590314-24; 10. 7590314-40; 11. 7590315-24; 12. 7590315-40; 13. 7590316-24; 14. 7590316-40; 15. 7590317-24; 16. 7590317-40; 17. 7590318-24-5; 18. 7590318-24-7; 19. 7590318-40-5; 20. 7590318-40-7; 21. 7590319-24-5; 22. 7590319-40-5; 23. 7590440; 24. 7590441; 25. 7590442; 26. 7590443; 27. 7590444; 28. 7590445; 29. 7590446; 30. 7590447; 31. 7590450-5; 32. 7590450-8
GN Otometrics Horskatten 9 Taastrup Denmark is recalling Gold Cup Leadwire and Snap Leadwire with the following component descriptions: 1. 24" Snap Leadw due to The electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm.
Recommended Action
Per FDA guidance
GN Otometrics notified customers on about 08/01/2019 via "URGENT FIELD SAFETY NOTICE (FSN) MEDICAL DEVICE RECALL" letter. The notification informed customers that the electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm and there is potential risk to healthcare professionals or patients of exposure to electrical shock form the electrode connector. Instructions included to identify, cease use and quarantine the affected electrode leads, ensure all personnel are notified who need to be aware, notify customers if the affected electrode leads have been further distributed, and arrange for the return of affected electrode leads to Natus Medical, 3150 Pleasant View Lane, Middleton, WI 53562. Customers were also instructed to complete and return the Customer Reply Form via email to otochartrLeads.US@natus.com or fax to 1-847-557-1308.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026