Electrical Meters (Greenlee) – Testing Failure (2002)
Check Your Product
According to the U.S. Consumer Product Safety Commission (CPSC)
Product
Greenlee CM-700 and CM-750 electrical meters
Brand
Greenlee Textron, Inc. (Greenlee), of Rockford, Ill.
Lot Codes / Batch Numbers
Not specified in CPSC notice. Check official source for details.
Product Images
Source: U.S. Consumer Product Safety Commission
Greenlee Textron, Inc. (Greenlee), of Rockford, Ill. is recalling Greenlee CM-700 and CM-750 electrical meters. The hazard is: An incompatible grommet, located in the battery compartment to protect internal wires, can cause the. Based on CPSC recall notice.
Summary derived from CPSC notice
The following details were not provided by CPSC: UPC codes.
Reason for Recall
As stated by CPSC
This recall involves certain CM-700 and CM-750 electrical meters. The 8-inch meters are dark green with bright yellow features. The model number and the words "Greenlee Test Instruments" are printed across the front of the meter. A silver plate on the back of the meter displays the serial number (S/N). Serial numbers included in the recall range from 0203540001 through 0203540650. The electrical meters were made in Taiwan. CM-700 and CM-750 meters outside the above serial number range are not included in this recall.
Reported concern (CPSC)
An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccurately indicates zero voltage or current creates the potential for electric shock or an electrocution hazard from the equipment being tested.
Recommended Action
Per CPSC guidance
Consumers should stop using the recalled meters immediately and contact Greenlee to receive a replacement meter.
Verify this information on the official source
For complete details and official instructions, check the U.S. Consumer Product Safety Commission (CPSC) directly.
View official CPSC recall noticePage updated: Jan 6, 2026