Guidant Corporation VIRTUS PLUS II (model nos. 1380 and 1480) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . VIRTUS PLUS II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy and providing various levels of therapeutic and diagnostic functionality. Two sensors are available; these adapt the pacing rate to the patient's changing metabolic demand. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VIRTUS PLUS II (model nos. 1380 and 1480) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . VIRTUS PLUS II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy and providing various levels of therapeutic and diagnostic functionality. Two sensors are available; these adapt the pacing rate to the patient's changing metabolic demand.
Brand
Guidant Corporation
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Affected devices may be contained in the following serial number ranges. Model 1380: serial numbers 100000 thru 100044. Model 1480: Serial numbers 500005 thru 500072.
Guidant Corporation is recalling VIRTUS PLUS II (model nos. 1380 and 1480) pacemaker. Sterilized with gaseous ethylene oxide. Gui due to A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within th. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
Recommended Action
Per FDA guidance
INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026