Gyrus Medical, Inc Gyrus PlasmaKinetic Super Pulse Generator is a micro-computer based radio frequency generator with electrical and software components made of biocompatible materials for medical applications. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Gyrus PlasmaKinetic Super Pulse Generator is a micro-computer based radio frequency generator with electrical and software components made of biocompatible materials for medical applications.
Brand
Gyrus Medical, Inc
Lot Codes / Batch Numbers
Model 744000
Products Sold
Model 744000
Gyrus Medical, Inc is recalling Gyrus PlasmaKinetic Super Pulse Generator is a micro-computer based radio frequency generator with e due to The generator may malfunction when used in close proximity to a monopolar generator.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The generator may malfunction when used in close proximity to a monopolar generator.
Recommended Action
Per FDA guidance
On December 05, 2003, the firm sent recall notices to their affected customers by Certified Mail. The customers were notified that the firm would call them to arrange for pick up and correction of the device.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026