Hamilton Medical Inc Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc., 4990 Energy Way,Reno, NV 89502-Serial Numbers 2975 to 5360, Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc., 4990 Energy Way,Reno, NV 89502-Serial Numbers 2975 to 5360,
Brand
Hamilton Medical Inc
Lot Codes / Batch Numbers
Serial Numbers 2975 to 5360
Products Sold
Serial Numbers 2975 to 5360
Hamilton Medical Inc is recalling Hamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical due to ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
Recommended Action
Per FDA guidance
On May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.