Hamilton Medical, Inc. Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Brand
Hamilton Medical, Inc.
Lot Codes / Batch Numbers
UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Products Sold
UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Hamilton Medical, Inc. is recalling Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator due to Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the bo. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Recommended Action
Per FDA guidance
On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.