Heart Sync, Inc. ADULT/CHILD Radiotranslucent Electrode, Part number T100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use. Recall
Class I: A situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death
Class I recall indicates potential for serious health consequences.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ADULT/CHILD Radiotranslucent Electrode, Part number T100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.
Brand
Heart Sync, Inc.
Lot Codes / Batch Numbers
All Lots. Catalog / Part Number T100AC-PHILIPS
Products Sold
All Lots. Catalog / Part Number T100AC-PHILIPS
Heart Sync, Inc. is recalling ADULT/CHILD Radiotranslucent Electrode, Part number T100AC-PHILIPS, Rx ONLY, For use with Philips He due to Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-c. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.