Heartsine Technologies Inc HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003.
Brand
Heartsine Technologies Inc
Lot Codes / Batch Numbers
Serial numbers within range of 0000101 and 00003986
Products Sold
Serial numbers within range of 0000101 and 00003986
Heartsine Technologies Inc is recalling HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003. due to Immediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Immediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts blinking indicating a fault condition.
Recommended Action
Per FDA guidance
AED software revision 1.0.20 and higher incorporate software enhancements to make the AED more forgiving for extreme variations in user technique. Firm will provide distributors with field correction notice dated 10/5/05. Customers to be notified by mid-October 2005.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026