Hero Model 100 (Hero Health) – Dose Dispensing Error (2021)
Software error can cause improper dose recording and potential medication mistakes.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management app.
Brand
Hero Health
Lot Codes / Batch Numbers
Model 100 (H-100) Catalog Number: F00002 Hero 100 V1.4
Products Sold
Model 100 (H-100) Catalog Number: F00002 Hero 100 V1.4
Hero Health is recalling The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication managem due to Due to dispenser prompting for a missed dose that had already been dispensed. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to dispenser prompting for a missed dose that had already been dispensed
Recommended Action
Per FDA guidance
On 07/02/2021, the firm initially sent out an "URGENT NOTICE" via email notifying customers of an product update correction. Customers were informed that the firm had become aware that a recent firmware update (v2.4.12) that users have experienced an error, which caused improper recording of a dispensed dose as having not been dispensed. This error may prompt a user to dispense the same dose again, and the mobile app will indicate that the previously dispensed dose has not been taken yet. Customers are instructed to: -If the pill dispenser prompts you to re-dispense a scheduled dose which has already been dispensed today, select Skip. -Use extra caution when dispensing doses prior to the scheduled firmware update and verify that the pill dispenser has a working internet connection. This will ensure that the firmware update (v2.4.14) will be automatically applied on all potentially affected online pill dispensers (at approximately 2:00 AM on July 3, 2021) to correct the issue. -To manually update firmware immediately, follow these steps: 1) Ensure the Pill Dispenser is connected to WiFi (Check for the WiFi icon on the main screen). 2) Go to the Device menu, by pressing the Center button on the main screen. 3) Scroll down and select Device Settings . 4) Scroll down and select Firmware Update. 5) Select I m Ready. The update could take several minutes. It is recommended to run the firmware update at least 30 minutes before a scheduled dose. -If the pill dispenser is not connected to WiFi, the firmware update cannot be completed. In this event, contact the Recalling Firm's Member support for assistance before continuing to use the pill dispenser to ensure the update is applied as soon as possible For question or further assistance, contact the Recalling firm at 855-855-9962 or via email at hello@herohealth.com (24/7).
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026