Hill-Rom, Inc. (1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR CARRIER SLIM/RAISED, LR 3101705 TRAV RAIL CARR ESYST WIDE 3102517 TR CARR. NORMAL/LOWER. LR 3102506 TR CARRIER WIDE/RAISED ML 3102519 TR CARR. LOW.LR 30MM/PAIR 3102511 TR CARRIER NORMAL, LR 3102531 TRAVERSE RL CARRIER STD/RAI Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR CARRIER SLIM/RAISED, LR 3101705 TRAV RAIL CARR ESYST WIDE 3102517 TR CARR. NORMAL/LOWER. LR 3102506 TR CARRIER WIDE/RAISED ML 3102519 TR CARR. LOW.LR 30MM/PAIR 3102511 TR CARRIER NORMAL, LR 3102531 TRAVERSE RL CARRIER STD/RAI
Brand
Hill-Rom, Inc.
Lot Codes / Batch Numbers
natural 3122007NA Likorall 242 ES R2R, natural 3122008 Likorall 242 ES R2R, white 3122008NA Likorall 242 ES R2R, white 3122011 Likorall 242 S R2R, natural 3122011NA Likorall 242 S R2R, natural 3122012 Likorall 242 S R2R, white 3122012NA Likorall 242 S R2R, white 3123001 Likorall 243 ES, natural 3123001NA Likorall 243 ES, natural 3123002 Likorall 243 ES, white 3123002NA Likorall 243 ES, white 3122501 Likorall 250 ES, natural 3122501NA Likorall 250 ES, natural 3122502 Likorall 250 ES, white 3122502NA Likorall 250 ES, white 3124050 Likorall 250 S, IRC, natural
Products Sold
Distributed between 04JAN2021 and 24MAR2021. P/N Product name 3130001 Multirall 200 3301030 LikoGuard L 3301040 LikoGuard XL 3121001 Likorall 200 3122005 Likorall 242 ES natural 3122005NA Likorall 242 ES natural 3122006 Likorall 242 ES white 3122006NA Likorall 242 ES white 3122009 Likorall 242 S natural 3122009NA Likorall 242 S natural 3122010 Likorall 242 S white 3122010NA Likorall 242 S white 3122007 Likorall 242 ES R2R, natural 3122007NA Likorall 242 ES R2R, natural 3122008 Likorall 242 ES R2R, white 3122008NA Likorall 242 ES R2R, white 3122011 Likorall 242 S R2R, natural 3122011NA Likorall 242 S R2R, natural 3122012 Likorall 242 S R2R, white 3122012NA Likorall 242 S R2R, white 3123001 Likorall 243 ES, natural 3123001NA Likorall 243 ES, natural 3123002 Likorall 243 ES, white 3123002NA Likorall 243 ES, white 3122501 Likorall 250 ES, natural 3122501NA Likorall 250 ES, natural 3122502 Likorall 250 ES, white 3122502NA Likorall 250 ES, white 3124050 Likorall 250 S, IRC, natural
Hill-Rom, Inc. is recalling (1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail C due to Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are u. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients
Recommended Action
Per FDA guidance
HillRom issued Urgent Medical Device Correction letter(FA-2021-11-004-LUL-002) on the 14th of January via UPS Ground service. Letter states reason for recall, health risk and action to take: Actions to be taken by Customer: Please identify if you have the potentially affected product in your facility (See affected device listing above). Do not use your traverse carrier until inspected by a Hillrom Technician. Return the response form below within two weeks. Upon receipt of the response form, Hillrom will contact you to schedule a visit to inspect your carriers. Should incorrect, non-conforming screws be present, these will be exchanged by an approved Hillrom technician. Action to be taken by the Distributor: Please share this Field Safety Notice with your end users and complete the attached response form and return to hillromLUL002@sedgwick.com within two weeks. Hillrom will provide replacement screws for Distributors to facilitate exchange among your end users. Action to be taken by Hillrom: Hillrom will arrange a site visit to inspect the potentially affected units. If short screws are identified, they will be exchanged with those of correct length. Contact reference person If you have any questions regarding this safety notice, please contact Hillrom Technical Support at 800-445-3720 or technical.support@hillrom.com .
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026