Hill-Rom, Inc. Hill-Rom 100 Low Bed; Model No. P3930. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hill-Rom 100 Low Bed; Model No. P3930.
Brand
Hill-Rom, Inc.
Lot Codes / Batch Numbers
All units.
Products Sold
All units.
Hill-Rom, Inc. is recalling Hill-Rom 100 Low Bed; Model No. P3930. due to The upper deck may collapse to its lowest position and the caster brakes may fracture if the bed is moved while the brakes are locked.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The upper deck may collapse to its lowest position and the caster brakes may fracture if the bed is moved while the brakes are locked.
Recommended Action
Per FDA guidance
Consignees were notified via letter dated April 8, 2008 to lower these beds to their lowest position, to remove the beds from use if they move more than 1/2 inch with the brakes locked and that the firm will repair these beds when fixes become available. If you have questions regarding this recall, contact Hill-Rom Technical Support at 800-445-3730.