Hill-Rom, Inc. Hill-Rom Affinity Four Birthing Bed, Model P3700B. Note: Not distributed in the U.S. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hill-Rom Affinity Four Birthing Bed, Model P3700B. Note: Not distributed in the U.S.
Brand
Hill-Rom, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
All units. (**Affinity beds manufactured with serial number I290AA7395 or greater are not impacted by this correction.)
Hill-Rom, Inc. is recalling Hill-Rom Affinity Four Birthing Bed, Model P3700B. Note: Not distributed in the U.S. due to Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear poin. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated.
Recommended Action
Per FDA guidance
Consignees were notified of the problem via an Urgent Medical Device Correction letter dated 11/21/07. Users were informed that they would be receiving a free upgrade kit for all of their Affinity Birthing Beds. The Kit includes two plugs per bed, an insertion tool, and full written instructions. An instruction video is available at www.Hill-Rom.com.