Hill-Rom, Inc. Hill-Rom Affinity Three Birthing Bed, Model P3700. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hill-Rom Affinity Three Birthing Bed, Model P3700.
Brand
Hill-Rom, Inc.
Lot Codes / Batch Numbers
All units distributed after October 2000 and equipped with the 12VAC accessory outlet option.
Products Sold
All units distributed after October 2000 and equipped with the 12VAC accessory outlet option.
Hill-Rom, Inc. is recalling Hill-Rom Affinity Three Birthing Bed, Model P3700. due to The cable on the auxiliary outlet may become pinched, which may result in an electrical short with melting of the plastic transformer and the emission. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The cable on the auxiliary outlet may become pinched, which may result in an electrical short with melting of the plastic transformer and the emission of smoke.
Recommended Action
Per FDA guidance
Customers were notified via letter dated 8/18/06 and instructed to either correct the problem themselves with the instructions provided or to contact the firm requesting a visit from the recalling firm to correct their beds.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026