Hologic, Inc Panther Fusion Tube Tray, REF PRD-04000, PN: FA-15004, Cont. 18 Tube Trays - Product Usage: an ancillary consumable to the Panther Fusion System that is used to contain the PCR and RT-PCR steps of the assay processing. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Panther Fusion Tube Tray, REF PRD-04000, PN: FA-15004, Cont. 18 Tube Trays - Product Usage: an ancillary consumable to the Panther Fusion System that is used to contain the PCR and RT-PCR steps of the assay processing.
Brand
Hologic, Inc
Lot Codes / Batch Numbers
Part No. (PN) FAB-15004 (Box Lot Number)/Catalog No PRD-0400 (Box/Tray Lot Number) 285762 / 647288, 285765 / 647288, 286025 / 648958, 286026 / 648958, 286028 / 648959, 286852 / 650655, 287410 / 650656, 289297 / 648959, 290057 / 647288, 290068/654833, 290409 / 654833, 291425 / 657545
Products Sold
Part No. (PN) FAB-15004 (Box Lot Number)/Catalog No PRD-0400 (Box/Tray Lot Number) 285762 / 647288, 285765 / 647288, 286025 / 648958, 286026 / 648958, 286028 / 648959, 286852 / 650655, 287410 / 650656, 289297 / 648959, 290057 / 647288, 290068/654833, 290409 / 654833, 291425 / 657545
Hologic, Inc is recalling Panther Fusion Tube Tray, REF PRD-04000, PN: FA-15004, Cont. 18 Tube Trays - Product Usage: an ancil due to There is a potential for tube trays to leak which could potentially result in false positive results, false negative results or invalid patient resul. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential for tube trays to leak which could potentially result in false positive results, false negative results or invalid patient results.
Recommended Action
Per FDA guidance
On February 26, 2021, Hologic, issued "Urgent: Medical Device Recall Notification" to affected consignees via FedEx. In addition, to informing consignees about the recall, they asked customers to take the following actions: Laboratory Action: A. Please immediately discontinue use of the Panther Fusion Tube Tray lots listed in Table 1. B. Please immediately check your inventory and segregate the Panther Fusion Tube Tray lots listed in Table 1. C. The number of tube tray boxes shipped to your site is identified on the accompanying document. (GBS customers will receive a blank form and will need to fill in lot number and quantity.) - Please fill out the documents completely, including the number of boxes of the specified lots remaining in inventory (whether unopened or partial boxes). - After completing the document, destroy the segregated inventory. - Sign the document to signify compliance and return it as instructed on the document. - Please do this even if you do not have any of the affected trays remaining in your inventory. D. Hologic will ship replacement product to your site based on the number of boxes of the specified lots you list as remaining in your inventory on the document you return. E. Please ensure that appropriate laboratory personnel are notified of the information herein. F. If you have any questions or concerns about this notification or about the replacement product, please contact Hologic Technical Support at 1888-484-474- or molecularsupport@hologic.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.