Home Access Health Corp (HAHC) Home Access-Hepatitis C Check; At Home Telemedicine Test Service for Hepatitis C ; each kit is composed of an outer box, a shrink-wrapped inner box that contains the blood collection card with an unique PIN, the instructions for use and ''Frequently Asked Questions'', a polybag component that includes two lancets, one gauze pad, one adhesive bandage and one alcohol prep pad, a specimen return envelope and a prepaid return mailer either designated as U.S. first class or FedEx; Home Access Health Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Home Access-Hepatitis C Check; At Home Telemedicine Test Service for Hepatitis C ; each kit is composed of an outer box, a shrink-wrapped inner box that contains the blood collection card with an unique PIN, the instructions for use and ''Frequently Asked Questions'', a polybag component that includes two lancets, one gauze pad, one adhesive bandage and one alcohol prep pad, a specimen return envelope and a prepaid return mailer either designated as U.S. first class or FedEx; Home Access Health
Brand
Home Access Health Corp (HAHC)
Lot Codes / Batch Numbers
UPC 0 83170 61000 2: lot 042715, exp May 2007, lot 042716, exp June 2007, lot 042717, exp June 2007, lot 042718, exp June 2007, lot 042719, exp August 2007, lot 042720, exp August 2007, lot 042721, exp August 2007, lot 042722, exp August 2007, lot 042723, exp October 2007, lot 042724, exp October 2007, lot 042725, exp October 2007, lot 042726, exp October 2007, lot 042727, exp October 2007, lot 052703, exp November 2007, lot 052704, exp November 2007, lot 052705, exp November 2007, lot 052706, exp November 2007, lot 052708, exp November 2007, lot 052709, exp January 2008
Products Sold
UPC 0 83170 61000 2: lot 042715, exp May 2007; lot 042716, exp June 2007; lot 042717, exp June 2007; lot 042718, exp June 2007; lot 042719, exp August 2007; lot 042720, exp August 2007; lot 042721, exp August 2007; lot 042722, exp August 2007; lot 042723, exp October 2007; lot 042724, exp October 2007; lot 042725, exp October 2007; lot 042726, exp October 2007; lot 042727, exp October 2007; lot 052703, exp November 2007; lot 052704, exp November 2007; lot 052705, exp November 2007; lot 052706, exp November 2007; lot 052708, exp November 2007; lot 052709, exp January 2008
Home Access Health Corp (HAHC) is recalling Home Access-Hepatitis C Check; At Home Telemedicine Test Service for Hepatitis C ; each kit is comp due to Wrong Expiration Date; The kits were labeled with Expiration/Use By dates; that exceed the expiration dates of the sterile safety lancet component of . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Wrong Expiration Date; The kits were labeled with Expiration/Use By dates; that exceed the expiration dates of the sterile safety lancet component of the kit.
Recommended Action
Per FDA guidance
Recall letters dated 10/6/06 were sent to the direct accounts via FedEx on the same date. The accounts were informed of incorrect Use By dates on the kits based on the lancet component of the kit, and that the lancet may not be sterile as of the printed Use By date on the kit. The letters listed the specific products and affected lot numbers shipped to the individual account, and requested them to examine their inventory for the affected product and return it via UPS to Home Access for replacement. No test results will be provided to any customer who uses the affected product after 10/31/06. Any questions were directed to 1-800-871-6046.