Hospira Inc. LifeShield Latex-Free HEMA Blood PlumSet, Nonvented, 110 Inch with 210 Micron Filter, Prepierced Injection Site and OPTION-LOK, Dual Channel, Capped Secondary Port; a sterile Rx blood administration set for use with the Plum Series Infuser; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; list number 11235-03 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
LifeShield Latex-Free HEMA Blood PlumSet, Nonvented, 110 Inch with 210 Micron Filter, Prepierced Injection Site and OPTION-LOK, Dual Channel, Capped Secondary Port; a sterile Rx blood administration set for use with the Plum Series Infuser; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; list number 11235-03
Brand
Hospira Inc.
Lot Codes / Batch Numbers
lot 25-086-5H
Products Sold
lot 25-086-5H
Hospira Inc. is recalling LifeShield Latex-Free HEMA Blood PlumSet, Nonvented, 110 Inch with 210 Micron Filter, Prepierced Inj due to Inlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Inlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.
Recommended Action
Per FDA guidance
Hospira sent product recall letters dated 5/6/05 to the accounts via Federal Express overnight letter on the same date. The letters informed the accounts of the possiblitiy of the inlet and outlet tubing being reversed on the cassette, which would allow the patient''s blood to be drawn up the tubing. The accounts were requested to examine their inventory for the affected lot, quarantine any found and call Hospira Customer Care at 1-877-946-7747 to arrange for return of the product for replacement.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.