In2Bones, SAS 28 Chemin du petit Bois Ecully France I.B.S 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
I.B.S 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175
Brand
In2Bones, SAS 28 Chemin du petit Bois Ecully France
Lot Codes / Batch Numbers
UDI/DI 3760225710685, Batch Number 2306007, exp. 31/MAY/2028
Products Sold
UDI/DI 3760225710685, Batch Number 2306007, exp. 31/MAY/2028
In2Bones, SAS 28 Chemin du petit Bois Ecully France is recalling I.B.S 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175 due to This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.
Recommended Action
Per FDA guidance
The firm notified its US consignee via email on 02/26/2024. The email identified the products begin recalled and attached the notice the consignee was to use to notify their customers. That notice, the FIELD CORRECTIVE ACTION, MEDICAL DEVICE BATCH RECALL - EXTENSION, was issued on 01/26/2024, via email, to the US consignees which explained the issue with the device, associated risks, and requested the return of the affected products.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026