Inlightened Partners LLC inLight Medical LED Pads used with Polychromatic light technology system Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
inLight Medical LED Pads used with Polychromatic light technology system
Brand
Inlightened Partners LLC
Lot Codes / Batch Numbers
All serial numbers with cord connectors that have v-shaped notches
Products Sold
All serial numbers with cord connectors that have v-shaped notches
Inlightened Partners LLC is recalling inLight Medical LED Pads used with Polychromatic light technology system due to LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.
Recommended Action
Per FDA guidance
On 6/5/2019, Medical Device Recall Letters were emailed to customers. Customers were instructed to examine the connector on the cord attached to their pad and compare it to photos provided in the Medical Device Recall Letter. The connectors for recalled product contain a v-shaped notch. Customers in possession of a pad with the connector matching the photos were requested to complete an Acknowledgement and Receipt Form. Customers with questions were asked to call Customer Service at 1-800-455-4116.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026