Inpeco S.A. Via San Gottardo 10 Lugano Switzerland FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Brand
Inpeco S.A. Via San Gottardo 10 Lugano Switzerland
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
The impacted modules are the Aliquoter Modules (lnpeco Part Number FLX-212) with the firmware versions listed below or higher: - AQMb_3-3-0.H86 - AQMa_3-1-1-8.H86 and AQMb_3-1-1-8.H86 - xAQMb_1-1-0.elf; System Serial Numbers: FLX.008 FLX.0118 FLX.064 FLX.101 FLX.0116 FLX.0125 FLX.080 FLX.0206 FLX.0161 FLX.0152 FLX.012 FLX.0121 FLX.0155 FLX.033 FLX.047 FLX.0127 FLX.0128 FLX.0129 FLX.074 FLX.076 FLX.0113 FLX.0138 FLX.014 FLX.015 FLX.0164 FLX.051 FLX.0106 FLX.0158 FLX.0196 FLX.072 FLX.0107 FLX.0189 FLX.0126 FLX.0165 FLX.0135 FLX.017 FLX.0114 FLX.022 FLX.0119 FLX.0105 FLX.029 FLX.0115 FLX.031 FLX.032 FLX.0141 FLX.037 FLX.020 FLX.039 FLX.043 FLX.102 FLX.046 FLX.0143 FLX.050 FLX.0124 FLX.0131 FLX.056 FLX.060 FLX.0136 FLX.0140 FLX.018 FLX.021 FLX.023 FLX.030 FLX.044 FLX.052 FLX.082 FLX.083 FLX.093 FLX.071 FLX.077 TSH.003 FLX.0146 FLX.027 FLX.038 FLX.075 FLX.0160 FLX.0153 FLX.0154 FLX.0157 FLX.0166 FLX.0167 FLX.0170 FLX.0174 FLX.0179 FLX.0180 FLX.0182 FLX.0194 FLX.0183 FLX.0048 FLX.0049 FLX.0109 FLX.0145 FLX.0148
Inpeco S.A. Via San Gottardo 10 Lugano Switzerland is recalling FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Acc due to When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first em. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
Recommended Action
Per FDA guidance
The firm, inpeco, sent "Urgent: Medical Device Correction" letters dated 7/14/20 to its customers. The letter described the product, problem and actions to be taken. Action to be taken by the user: Be aware that the Secondary Sample Tube flagged with error 2132 or 1442 may be diluted: discard this Secondary Sample Tube or manage it according to your laboratory guidelines. Your service provider will contact you to schedule the upgrade of the software driver and language.ini with the new error message and error recovery displayed if error 2132 or 1442 occurs. Until the service visit please maintain awareness on this notice. Please transfer this notice to whom it might concern. Please complete and return the Field Safety Notice Receipt Confirmation form attached to this letter within 15 days directly to the email address specified in the email communication. Contact reference person: For any clarification you may need, do not hesitate to contact: Eva Balzarotti - Regulatory Affairs Manager E-mail: Regulatory.Affairs@inpeco.com Phone: (+41) 919118 224
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026