Integra LifeSciences Corp. Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332
Brand
Integra LifeSciences Corp.
Lot Codes / Batch Numbers
Lot #AA1901
Products Sold
Lot #AA1901
Integra LifeSciences Corp. is recalling Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clam due to A small crack at the box lock corner may after multiple cleaning and sterilization cycles produce an increased risk of corrosion, weakening the instru. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A small crack at the box lock corner may after multiple cleaning and sterilization cycles produce an increased risk of corrosion, weakening the instrument and resulting in premature need for replacement and/or breakage of the instrument
Recommended Action
Per FDA guidance
Integra LifeSciences notified consignees (Hospitals, Distributors) by letter delivered via FedEx on August 7, 2019. Letter states reason for recall, health risk and action to take: 1. Review your inventory and determine if you have any impacted lots listed above.2. If you do have product with an affected Lot Number, stop using the product immediately and remove the product from service. 3. Complete the attached form. If you do not have affected product, check the box, I do not have any product with the lot number. 4. If you do have affected product, check the box I do have product(s) with the lot number. Record the total quantity of affected product you have.5. Complete the other information and Return the completed Acknowledgement and Return Form by email or fax indicating your receipt and review of this notification. Keep a copy of the form for your records. Questions contact Customer Service Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com Outside USA: 781-565-1401.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.